Multicentre dosimetry audit in the European randomized phase III study of neoadjuvant chemoradiotherapy using proton versus photon radiotherapy in locally advanced oesophageal cancer.
Mai L Ehmsen, Camilla S Byskov, Hanna R Mortensen, Rebecca Bütof, Richard Canters, Gilles Defraene, Karin Haustermans, Maria Fuglsang Jensen, Arturs Meijers, Christina T Muijs, Ditte S Møller, Marianne Nordsmark, Pieter Populaire, Gloria Vilches-Freixas, Lars Nyvang, Liliana Stolarczyk, Sebastian Makocki, Sara Broggi, Francesco Fracchiolla, Andrea Martignano, Kenneth Poels, Séverine Rossomme, Stefan Vasiliniuc, Lone Hoffmann
Phys Imaging Radiat Oncol 38 100938
摘要
摘要
This study presents the results of a multicentre external dosimetry audit conducted as part of the PROTECT trial, involving nine European radiotherapy centres delivering proton and photon treatments. A dosimetry equipment vendor performed beam output audits at participating sites, verifying the absorbed dose-to-water measurements. Measurements used reference conditions based on the IAEA TRS-398, with ionization chambers traceable to primary standards. The audit compared local centre measurements to independent assessments, finding all within the acceptable 3% agreement threshold. The study is part of the successful implementation of a comprehensive, multicentre QA programme in the PROTECT trial.
KM 提取结果
0 个 Figure · 0 个 KM 面板